Required with Smart Reader 5 Data Logger
>> TrendReader 2 Software
Required for Ethernet Connectivity
>> TrendReader 2 Software with USB Cable Kit (#TR2-USB)
NOTE: NIST Calibration is available.
Large Quantity Discounts Are Available!
| Description | Part Number | |||
| SmartReader 5 Three Channel Temperature (thermocouple) logger | SR-005 | |||
| Software Kit - 1 Required | ||||
| TrendReader 2 Software with RS-232 Cable (IC-101) and Manual for Smart Reader series | TR2-RS232 | |||
| TrendReader 2 Software with USB Interface Cable (IC-102) and Manual for Smart Reader series | TR2-USB | |||
| TrendReader 2 Software CD Only (interface cable and manual not included), for SmartReader Plus, Smart Reader, IAQ and OWL Loggers | TR2-Standard | |||
| Standard NIST Calibration Service | ||||
| Standard NIST Temperature Calibration (1 Point) Note: 3 to 5 Day Lead Time for All Calibration Orders | CAL-NIST-TEMP | |||
| Optional Accessories | ||||
| USB Serial Adapter | CABLE-USB232 | |||
| ACR RS-232 Serial Interface Cable (Limited Quantity - 2 Available) | LIC-101--U-- | |||
| ACR RS-232 Serial Interface Cable | LIC-101 | |||
Specifications:
| General Specifications | |
|---|---|
| Channels | Three (one internal thermistor for cold-junction compensation and two external channels for narrow range and wide range thermocouples) |
| Memory | 32 KB (32,767 readings) |
| Size | 107 x 74 x 22 mm. (4.2” x 2.9” x 0.9”) |
| Weight | 110 grams (3.75 ounces) |
| Case Material | Noryl Plastic |
| External Connector | Removable screw-type terminal strip with Common (-) connection |
| Mounting | Magnetic backing or locking eyelet |
| Operating Limits | -40°C to 70°C (-40°F to 158°F) and 0 to 95% Relative Humidity (non-condensing) |
| Clock Accuracy | ±2 seconds per day |
| Battery | 3.6 volt Lithium, 1 Amp-Hour |
| Battery Life | 10 year warranty (under normal use), factory replaceable |
| Power Consumption | 5 to 10 microamps (continuous) |
| Sampling Methods | 1. Continuous (First-in, First-out) 2. Stop when full (Fill-then-stop) |
| Sampling Rates | User selectable rates from 8 seconds to once every 5 days |
| Resolution | 8 bit (1 part in 256) |
| Software Compatiblity | Windows 3.1, 95, 98, 2000, ME, NT, XP, Vista (32-bit), Windows 7 (32-bit and 64-bit) |
| Software Requirements | TrendReader 2 Software and Interface Cable |
| Certifications and Standards |
- Certified to CE standard EN50082-1: 1992 (European Generic Immunity) covering ESD, RFI and EFT/B |
| Resistance to X-Rays | - Tested for protection against a 160kV dose @ 5mA for 30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times that of an airport x-ray machine) - Tested for protection against Gamma Ray (equivalent to 0.137-1.38 mega volts) IR 192 - 28 curies @ 30 second exposure source to object distance 5” |
| Internal Temperature Sensor | |
|---|---|
| Type | Negative Temperature Coefficient Thermistor. 10,000 ohms @ 25°C (77°F) |
| Accuracy | ± 0.2°C, over the range of 0 to 70°C (±0.3°F over the range of 32 to 158°F) |
| Range | -40 to 70°C (-40 to 158°F) |
| Resolution | 0.4°C (0.7°F) at 25°C (77°F); better than 1 °C (1.8°F) between -25°C and 70°C (-13°F and 158°F); better than 2.0°C (3.6°F) between -40°C and -25°C (-40°F and -13°F) |
| Thermocouples Accepted | |||
|---|---|---|---|
| Type | Wide/Narrow | Range | Resolution |
| J | Channel #2 (Wide) | -50 to 600°C (-55 to 1100°F) | 5°C (9°F) |
| Channel #1 (Narrow) | -20 to 190°C (0 to 370°F) | 1.4°C (2.5°F) | |
| K | Channel #2 (Wide) | -100 to 900°C (-145 to 1650°F) | 6.7°C (12°F) |
| Channel #1 (Narrow) | -25 to 230°C (-10 to 440°F) | 1.7°C (3°F) | |
| S | Channel #2 (Wide) | 0 to 1760°C (32 to 3200°F) | 23°C (41.5°F) |
| Channel #1 (Narrow) | 0 to 960°C (32 to 1750°F) | 7°C (12.6°F) | |
| T | Channel #2 (Wide) | -200 to 400°C (-325 to 750°F) | 6.7°C (12°F) |
| Channel #1 (Narrow) | -35 to 200°C (-30 to 390°F) | 1.7°C (3°F) | |
| Accuracy: ± 1% of range + resolution | |||