>> Graphical LCD Display with Analysis Functions Shows Temperature,
Voltage and Current Readings
>> Sampling Rate of 4,000 per Second on 1 Channel
>> Stores up to 512,000 Total Samples/Readings (Temperature, Voltage, Current, Pulse or Frequency)
>> Internal 7.2 Volt Nickel Cadium Rechargeable Battery with 40 Hour Life Between Charges
>> Scale Readings, On Screen Text Editing of Collected Data and Keypad Setup
>> Includes DaqPRO Lab Software, Carrying Case, AC/DC Adpater, USB Cable, User Manual on CD and NIST Calibration Certificate
NOTE: Large Quantity Discounts Are Available!
Specifications
Channels | 8 Differential Inputs |
Temperature (Internal) | Type: Internal Sensor Range: -25°C to 70°C Resolution: 0.1°C (1 µV) Accuracy: ±0.5% FS |
Temperature (NTC Probe) |
NTC Type: 10/100KΩ Resistor Range: -25°C to 150°C Resolution: 0.05°C Accuracy: ±0.5% FS |
Temperature (PT-100 Probe) |
Channels: (8) PT-100 2-Wire or (4) PT-100 3-Wire Range: -200°C to 400°C Resolution: 0.1°C (7 mΩ) Accuracy: ±0.5% FS |
Thermocouple (Type J) |
Range: -200°C to 1200°C Resolution: 0.1°C (1 µV) Accuracy: ±0.5% FS Cold Junction Compensation Error: ±0.5°C |
Thermocouple (Type K) |
Range: -250°C to 1200°C Resolution: 0.1°C (1 µV) Accuracy: ±0.5% FS Cold Junction Compensation Error: ±0.5°C |
Thermocouple (Type T) |
Range: -200°C to 400°C Resolution: 0.1°C (1 µV) Accuracy: ±0.5% FS Cold Junction Compensation Error: ±0.5°C |
Current | Range: 0 to 24 mA Resolution: 0.47 µA Accuracy: ±0.5% FS Loop Impedance: 195 Ω |
Voltage (0 to 50 mVolts) |
Range: 0 to 50 mVolts Resolution: 3 µV Accuracy: ±0.5% FS |
Voltage (0 to 10 Volts) |
Range: 0 to 10 Volts Resolution: 200 µV Accuracy: ±0.5% FS Input Impedance: 125 kΩ |
Pulse Counter (Input 1 Only) |
Input Type: Optocoupler Input Range: 0to 65,000 Input Signal: 0 to 5 Volts Input Impedance: 470 Ω Bandwidth: 0 to 50 Hz |
Frequency Meter (Input 1 Only) |
Input Type: Optocoupler Input Range: 20 to 4,000 Hz Input Signal: 0 to 5 Volts Input Impedance: 470 Ω |
Analog to Digital | Noise: 30 µV rms Internal Linearity Error: ±0.08% of FSR Offset Error: 0.1% |
Open Collector Output (Output B) |
Maximum Current Sink: 50 mA (Fuse Protected) Input Impedance: 50 Ω Offset Error: 0.1% |
Data Storage Capacity | 512,000 Readings/Samples |
Sampling Rate | Variable, 1 sample/hour to 4,000 samples/sec, 1 channel Note: Number of Sensors in Use Limits Max. Sampling Rate. See User Manual (Section 1.2.10) for Details |
Sampling Resolution | 16 bits |
Power Supply | Internal Rechargeable 7.2V NiCa Battery Built in Battery Charger External 9V to 12V DC input |
Battery Life | 40 hours Between Charges |
Communication | USB 1.1 Compliant |
Housing | Plastic ABS Box |
Operating Range | Temperature: 0°C to 50°C |
Man Machine Interface | Full Keyboard Operation - enables manual programming of the logger Graphic LCD 64x128 pixels |
Software | DaqLab PC Suite for Windows (Included) |
OS Compatibility | Windows 95, 98, 2000, ME, XP and Vista |
Standards Compliance | CE, FCC |
Weight | 450g (15.87 oz.) |
Dimensions | 182 mm x 100 mm x 28 mm (7.16" x 3.93" x 1.10") |
DaqPRO 8-Channel Data Logger
|
Part #: DaqPRO-5300 | |
Standard NIST DC Voltage Calibration Service >> 1 Point (per Channel) Offset Calibration
>> Traceable to NIST >> Certificate Supplied >> Note: 3 to 5 Day Lead Time for All Calibration Orders |
Part #: CAL-NIST-DCVOLT |
|
Standard NIST 4-20mA Calibration Service >> 1 Point Offset Calibration
>> Traceable to NIST >> Certificate Supplied >> Note: 3 to 5 Day Lead Time for All Calibration Orders |
Part #: CAL-NIST-420MA |
Compatible Sensors & Probes
220/12V AC/DC Adapter >> Designed for Europe
>> Limited Quantity: 8 Available |
Part #: DT245--U-- |
Accessories
DatPass Administrative Software Package ( 21 CFR Part 11) with Dongle >> Used with Other Fourier Software Programs
>> Provides Compliance to FDA 21 CFR Part 11 Regulations |
Part #: SFTMCL025A | |
Solar Cell & Battery for DaqPRO >>
>> >> |
Part #: AC029 | |
DaqPRO Weather Box >>
>> >> |
Part #: 11460A | |
USB Communications Cable USB communication cable for MicroLog Cradle or DaqPRO |
Part #: DT180 | |
110/12V AC/DC Adapter >> Designed for US
|
Part #: DT246 | |
220/12V AC/DC Adapter >> Designed for UK
|
Part #: DT282 | |
220/12V AC/DC Adapter >> Designed for Europe
>> Limited Quantity: 8 Available |
Part #: DT245--U-- | |
220/12V AC/DC Adapter >> Designed for Europe
|
Part #: DT245 | |
2-Position Screw Terminal for DaqPro >>
>> >> >> |
Part #: ED-1960-ND | |
DigiKey Terminal Block Plug >> 4 Position
>> 3.81 mm |
Part #: ED1962--U-- |
Documentation
DaqPro and DaqLab User Manual
DaqPro Detailed Specifications
DaqPro Calibration Procedure & Guide
Working with 4-20mA Transducers
DatPass User Guide