Graphtec High Speed Isolated Data Logger - DATA LOGGER MALAYSIA OFFICIAL DISTRIBUTOR | DATA LOGGER MALAYSIA OFFICIAL DISTRIBUTOR

Graphtec High Speed Isolated Data Logger

 

 

The GL900 Data Logger is a high-speed 8 Channel recording device with isolated channels.The logger features a built-in web server, ftp server, LCD TFT Display, remote-monitoring, fast sampling rate of 10µs per channel and external storage using USB memory stick.

 

Included With Each Logger 
>> User Manual & Quick Start Guide 
>> GL900-APS Software
>> 110 Volt AC Power Supply (Adapter) 
>> USB Cable

Optional Accessories Available
>> Humidity Sensor (#B-530)
>> 6 Hour Battery Pack (#B-517)
>> DC Power Cable (#B-514)

High Speed Isolated Data Logger Features

  • High-Speed Isolated 8 Channel Multifunction Data Logger
  • Simultaneous Measurements of Voltage, Temperature, Humidity, Pulse and Logic
  • Data Storage on USB Memory Sticks or Internal RAM for Samplings faster than 1 ms
  • Wide 500 Volt Range Enables 100 to 240 Volts AC Power Supply Voltage Waveform Measurements
  • Built-in 5.7" TFT LCD Display
  • Remote Monitoring with Built-in Ethernet Web and FTP Server
  • Includes Software, Power Supply and USB Cable
Description Part Number
Graphtec High Speed Isolated Data Logger GL900
Optional Thermocouples
Welded Tip Teflon Thermocouple w/ 72" Bare Wire Leads Welded Tip Teflon Thermocouple w/ 72" Bare Wire Leads
>> Operating Range: -75°C to 250°C
>> Leads: 72" Teflon Insulated Twisted Pair Bare Wire Leads
>> Available in Type J, K, T and N
Quantity Price Each
1 - 9  
10 - 25  
26+ $
TC Type:


Welded Tip Teflon Thermocouple w/ 72" Bare Wire Leads Welded Tip Teflon Thermocouple w/ 180" Bare Wire Leads
>> Operating Range: -75°C to 250°C
>> Leads: 180" Teflon Insulated Twisted Pair Bare Wire Leads
>> Available in Type J, K, T and N
Quantity Price Each
1 - 9  
10 - 25  
26+  
TC Type:

Optional Accessories
Graphtec Battery Pack B-517
Cable: 4x logic inputs, 4x alarm outputs, 2 meters B-513
DC Power cable, 2 meters, bare ends, for direct connection to DC power source B-514
Graphtec Humidity Sensor B-530
Graphtec BNC to BNC 6ft Cable RIC-112
Graphtec BNC to Banana 6ft Cable RIC-113
Graphtec BNC to Alligator Clips 6ft Cable RIC-114
Optional NIST Calibration
Standard NIST DC Voltage Calibration (1 Point) Note: 3 to 5 Day Lead Time for All Calibration Orders CAL-NIST-DCVOLT

High Speed Isolated Data Logger Specifications

Analog Inputs 8 channels
External Input/Output Trigger input (1 channel), Logic input (4 channels) or Pulse input (4 channels), Alarm output (4 channels)
Sampling Interval 10μs to 1 min
TIME/DIV 10 ms/DIV to 24 hour/DIV
Timer Functions Date and time, daily cycle, hourly cycle
Trigger Functions Type Start: Data capture starts when a trigger is activated;
Stop: Data capture stops when a trigger is activated
Condition Start: Off, Input signal level (analog, logic/pulse), External*1
Stop: Off, Input signal level (analog, logic/pulse), External *1, Scheduled time
Combination Input Signal Level: Level OR, Level AND, Edge OR, Edge AND
Mode H (Rising), L (Falling), Window In*2, Window Out*2
Alarm Setting Functions Rising, Falling, Window In*2, Window Out*2
Alarm Output *1 Number of Channels: 4, Open collector output (5V, 10 kΩ pull-up resistance)
Pulse Input*1, *3 RPM Mode 5 to 20 M RPM/F.S. (in steps of 1, 2, or 5)
Count Mode 5 to 20 M C/F.S. (in steps of 1, 2, or 5)
Inst. Mode 5 to 20 M C/F.S. (in steps of 1, 2, or 5)
Calculation Functions Statistical calculations *4: Average, Peak, Maximum, Minimum, RMS (2 calculations can be set simultaneously)
Other Functions Search function, annotation input function
PC Inteface Ethernet: 10BASE-T/100BASE-TX
USB:(High Speed) provided as standard
Ethernet Functions Web Server, FTP, NTP Client Functions
USB Function USB drive mode (File transfer and deletion from internal GL900 memory)
Internal Data Storage One million data points / Internal flash memory:Approx. 256 MB
External Data Storage USB memory slot (High speed supported) *5
Display Functions Waveforms + digital values, enlarged waveforms, digital values + calculation results, X-Y
Display 5.7" TFT color LCD
Operating Environment Temperature: 0°C to 40°C
Humidity: 5% to 85% RH (15°C to 35°C - Battery Power)
Withstand Voltage Between each input channel and GND: 1000 V p-p for one minute, between input terminals: 1000 Vp-p for one minute
Power Supply AC Adapter: 100 to 240 VAC, 50 to 60 Hz
DC Input: 8.5 to 24 Volts DC
Battery Pack: Optional (2 Battery Packs Required)
Power Consumption 28 VA
Dimensions 232 x 150 x 80 mm (W x H x D), approx.
Weight 1.1 kg (excluding AC adapter and battery)
Vibration-tested Conditions Equivalent to automobile parts Type 1 Category A Classification

Terminal block specifications

Input Channels Fixed to 8 channels
Input Terminal Type Voltage: BNC connector
Temperature: M3 screw type terminal board *7
Input Method All channels Isolated, Imbalanced Input, Simultaneous Sampling of all Channels
Measurement Ranges Voltage: 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100, 200, 500 V F.S., 1-5 V F.S.
Thermocouples: K, J, E, T, R, S, B, N, W (WRe5-26)
Humidity: 0 to 100% (voltage 0 V to 1 V scaling conversion) * with Optional B-530 Humidity Sensor
Input Filter Off, Line, 5 Hz, 50 Hz, 500 Hz
A/D Converter 16 bits (14 bits internally acknowledged)
Maximum Permissible Input Voltage Between Input Channel + and - Terminals:
20 mv to 1 V: 30 Vp-p
2 V to 500 V: 500 Vp-p
Between Input Channel Terminals: 60 Vp-p
Between Input Channel Terminal and GND Terminal: 60 Vp-p
Withstand Voltage Between Input Channel Terminal and GND Terminal: 1 minute at 1000 Vp-p
Between Input Channel Terminals: 1 minute at 1000 Vp-p

Notes:
*1 Logic Alarm Cable (#B-513) is Required
*2 Cannot be Set for Logic Input
*3 Maximum Input Frequency: 50 kHz, Maximum Number of Counts: 15 M C
*4 In Real Time or when Between Cursors has been Specified (During Replay)
*5 1 File = 2 Gbytes (Depends on the USB Memory Stick Used)
*6 Please Install Two Battery Packs
*7 Connections made to both the BNC terminal and M3 Screw Terminal For The Same Channel
*8 Thermocouple Diameters T:0.32φ, others: 0.65φ
*9 Operating Temperature Range: -25°C to +80°C

Control Software Specifications

Supported OS Windows 2000, Windows XP, Windows Vista (32-bit and 64-bit Versions)
Functions GL900 control, real-time data capture, data conversion
Setting Range Amp settings, data capture settings, trigger settings, alarm settings, other
Display Information Analog waveforms, logic waveforms, pulse waveforms, digital values
File Conversion Data between cursors, All data
2-Screen Function (Zoom) Display of current and past data
Display of Statistics and History Display of maximum, minimum, and average values

Application Notes

Understanding the Differences Between State, Event and Pulse Datalogging. (.pdf 300KB)