SmartReader Plus 4 Differential Data Logger

 

The ACR Low Pressure Differential SmartReader Plus 4 data logger is designed to be used in applications where precise and very low differential pressure measurements are required.

Required with Temperature Data Logger
>> TrendReader 2 software

Required for Ethernet Connectivity 
>> TrendReader 2 Software with USB Cable Kit (#TR2-USB)
>> Read the Networking and Remote Communications Guide

NOTE: NIST Calibration is available.
Large Quantity Discounts Are Available!

SmartReader Plus 4 Logger Features

  • Extremely Accurate
  • 10 Year Battery Life
  • Data Storage Capacity of 21,500 Readings
  • 2 Channels

SmartReader Plus 4 Logger Specifications

General Specifications
Channels Two (one for the internal temperature sensor and one for the internal pressure sensor)
Memory Size 32 KB (21,500 readings)
Size 107 x 74 x 22 mm. (4.2” x 2.9” x 0.9”)
Weight 110 grams (3.75 ounces)
Case Material Noryl Plastic
External Connector Removable screw-type terminal strip with Common (-) connection
Operating Limits -40°C to 70°C (-40°F to 158°F) and 0 to 95% Relative Humidity (non-condensing)
Clock Accuracy ±2 seconds per day
Battery 3.6 volt Lithium, 1 Amp-Hour
Battery Life 10 year warranty (under normal use), factory replaceable
Power Consumption 5 to 10 microamps (continuous)
Sampling Methods 1. Continuous (First-in, First-out – not available with sample rates faster than 8 seconds)
2. Stop when full (Fill-then-stop)
Sampling Rates User selectable rates from 25 per second to 1 reading every 8 hrs. (BP-101 Battery Pack or PS-201 Power Supply required for sample rates faster than 8 seconds)
Resolution 12 bit (1 part in 4,096)
Mounting Magnetic backing or locking eyelet
Software Compatiblity Windows 3.1, 95, 98, 2000, ME, NT, XP, Vista (32-bit), Windows 7 (32-bit and 64-bit)
Software Requirements TrendReader Software and Interface Kit
Product Approvals Meets Part 15 for Digital Devices of the Code of Federal Regulations of the Federal Communications Commission (FCC). Meets radiated and conducted emission require ments for the Interference Causing Equipment Standard for Digital Apparatus of Industry Canada (ICES-003). IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity, EFT/Burst and Radiated Emissions respectively for the Commission of the European Communities (CE).
Resistance to X-Rays Tested for protection against a 160kV dose @ 5mA for 30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times that of an airport x-ray machine). Tested for protection against Gamma Ray (equivalent to 0.137-1.38 mega volts) IR 192 - 28 curies @ 30 second exposure source to object distance 5”.
Pressure Sensor
Type Silicon piezoresistive strain gauge
Media Non-corrosive gases and dry air
Maximum Pressure Four times full scale
Pressure Sensor Ranges Available
Range Accuracy Resolution
± 0.5" WC (± 120 Pa) ± 0.05" WC (± 2.5 Pa) < 0.001" WC (0.25 Pa)
± 1" of WC (± 250 Pa) ± 0.05" WC (± 2.5 Pa) < 0.001" WC (0.25 Pa)
± 2" of WC (± 500 Pa) ± 0.05" WC (± 2.5 Pa) < 0.001" WC (0.25 Pa)
± 5" of WC (± 1.2 kPa) ± 0.1" WC (± 12.5 Pa) < 0.01" WC (2.5 Pa)
± 10" of WC (± 2.5 kPa) ± 0.1" WC (± 12.5 Pa) < 0.01" WC (2.5 Pa)
Internal Temperature Sensor
Type Negative Temperature Coefficient Thermistor.
10,000 ohms @ 25°C (77°F)
Accuracy ± 0.2°C, over the range of 0 to 70°C (±0.3°F over the range of 32 to 158°F)
Range -40 to 70°C (-40 to 158°F)
Resolution 0.03°C (0.05°F) at 25°C (77°F). Better than 0.07°C
(0.12°F) between -25°C (-13°F) and 70°C (158°F).
Better than 0.13°C (0.23°F) between -40°C (-40°F)
and -25°C (-13°F)

SmartReader Plus 4 Logger Application Notes (pdf)

Fast Sampling with Battery Pack or External 120 Volt Power Supply

For sampling rates faster than once every 8 seconds, the Smart Reader Plus data loggers require the use of an external 120 volt power supply (#PS-201) or fast sampling battery pack (#BP-101).

The fast sampling battery pack has replaceable "AAA" batteries and 8" (20cm) cable attached, with a male stereo jack on the end. The stereo jack is plugged into the communication port on the Smart Reader Plus data logger.
The 120 Volt Power Supply is required when network length is greater than 50' , more than 3 loggers are on the network, or for sampling rates greater than 8 seconds on the SmartReader Plus Loggers.

Description Part Number
SmartReader Plus 4 32KB, 2-channels: one differential pressure 2" WC, one temp (Limited Quantity 1 Available) SRP-004-LPD-2--U--
SmartReader Plus 4 32KB, 2-channels: one differential pressure 0.5" WC, one temp SRP-004-LPD-0-5
SmartReader Plus 4 32KB, 2-channels: one differential pressure 1" WC, one temp SRP-004-LPD-1
SmartReader Plus 4 32KB, 2-channels: one differential pressure 2" WC, one temp SRP-004-LPD-2
SmartReader Plus 4 32KB, 2-channels: one differential pressure 5" WC, one temp SRP-004-LPD-5
SmartReader Plus 4 32KB, 2-channels: one differential pressure 10" WC, one temp SRP-004-LPD-10
Software Kit - 1 Required
TrendReader 2 Software with RS-232 Cable (IC-101) and Manual for Smart Reader series TR2-RS232
TrendReader 2 Software with USB Interface Cable (IC-102) and Manual for Smart Reader series TR2-USB
TrendReader 2 Software CD Only (interface cable and manual not included), for SmartReader Plus, Smart Reader, IAQ and OWL Loggers TR2-Standard
Standard NIST Calibration Service
Standard NIST Temperature Calibration (1 Point) Note: 3 to 5 Day Lead Time for All Calibration Orders CAL-NIST-TEMP
Optional Accessories
Fast Sampling Battery Pack BP-101
Network/Fast Sampling Power Supply - 120V - North American Version PS-201-NA
Network/Fast Sampling Power Supply - 220V - European Version PS-201-EU
Extra RS-232 Interface Cable for SmartReader, SmartReader Plus and Nautilus loggers IC-101
USB Serial Adapter CABLE-USB232