8-Channel Temperature Data Logger - DATA LOGGER MALAYSIA OFFICIAL DISTRIBUTOR | DATA LOGGER MALAYSIA OFFICIAL DISTRIBUTOR

8-Channel Temperature Data Logger

 

 

The USB-5203 from Measurement Computing is an 8 channel temperature measurement device that will accept thermocouple, RTDs, thermistors and semiconductor sensors. The USB-5201 also has 8 independent temperature alarms. Each alarm controls an associated digital I/O channel as an alarm output.

The USB-5203 provides 2 integrated cold junction compensation (CJC) sensors for thermocouple measurements, and built-in current excitation sources for resistive sensor measurements. An open thermocouple detection feature lets you detect a broken thermocouple. An on-board microprocessor automatically linearizes the measurement data according to the sensor category.

NOTE: External AC power is required for data logging operations. Due to processing limitations, you cannot log data to the memory card when the USB-5201 is connected to your computer's active USB bus. When operating as a data logger, disconnect the USB cable from the computer, and connect the external power supply shipped with the device.

8 Channel Multi Sensor Data Logger includes:
>> Compact Flash Memory Card (512MB Min.)
>> 2 meter long USB Cable
>> 2.5 watt USB / AC Power adapter
>> Measurement Computing’s DAQ software suite that includes, TracerDAQ, The Universal Library, Comprehensive drivers for DASYLab and NI LabVIEW as well as InstaCal software utility for installation, calibration, and testing.

NOTE: Large Quantity Discounts Are Available!

Note: Any order for Measurement Computing Products to be Shipped Outside the US will require an "End User Export Agreement" form to be Completed and Returned Prior to Products being Shipped.

 

8-Channel Temperature Data Logger 8-Channel Temperature Data Logger
>> 8 External Input Channels
>> Supports Thermocouples, RTDs, Thermistors and Semiconductor Temperature Sensors
>> 8 digital I/O Bits with User-Configurable in Internal Memory for Alarming
>> Includes 512 MB CompactFlash½ Card, USB Power Adapter, US Plug and Software
Part #: USB-5203


TracerDAQ Pro Software TracerDAQ Pro Software
>> Strip Chart, Oscilloscope, Function Generator and Rate Generator Software
>> Create Charts and Export Data
Part #: TRACERDAQ-PRO

Das-Wizard Software for USB DAQ Devices Das-Wizard Software for USB DAQ Devices
>> DAQ Data to Excel Add-on Program
>> Places Measurements directly into an Excel Spreadsheet
>> Supports MCC DAQ Devices: PCI, USB, PC-Card, ISA and PC/104 Boards
>> No data conversion required. Measurements go directly to cells as volts, temperature or A/D counts
Part #: DAS-WIZARD

DASYLab Lite DAQ Software DASYLab Lite DAQ Software
>> Graphical Programming Software Package for Customized Applications Without Writing Code
>> Create Flow Charts Where Functions are Connected By Wires
>> Available in 4 Versions (Lite, Basic, Full and Pro)
>> Includes Hardware Drivers for Measurement Computing DAQ Devices
Part #: DASYLAB-LITE

DASYLab Lite DAQ Software DASYLab Basic DAQ Software
>> Graphical Programming Software Package for Customized Applications Without Writing Code
>> Create Flow Charts Where Functions are Connected By Wires
>> Available in 4 Versions (Lite, Basic, Full and Pro)
>> Includes Hardware Drivers for Measurement Computing DAQ Devices
Part #: DASYLAB-BASIC

DASYLab Lite DAQ Software DASYLab Full DAQ Software
>> Graphical Programming Software Package for Customized Applications Without Writing Code
>> Create Flow Charts Where Functions are Connected By Wires
>> Available in 4 Versions (Lite, Basic, Full and Pro)
>> Includes Hardware Drivers for Measurement Computing DAQ Devices
Part #: DASYLAB-FULL

DASYLab Lite DAQ Software DASYLab Pro DAQ Software
>> Graphical Programming Software Package for Customized Applications Without Writing Code
>> Create Flow Charts Where Functions are Connected By Wires
>> Available in 4 Versions (Lite, Basic, Full and Pro)
>> Includes Hardware Drivers for Measurement Computing DAQ Devices
Part #: DASYLAB-PRO

DASYLab Lite DAQ Software Additional DASYLab Runtime License
>> Graphical Programming Software Package for Customized Applications Without Writing Code
>> Create Flow Charts Where Functions are Connected By Wires
>> Available in 4 Versions (Lite, Basic, Full and Pro)
>> Includes Hardware Drivers for Measurement Computing DAQ Devices
Part #: DASYLAB-RUNTIME

 

Optional Accessories

 

Type K Pipe Clamp Probe Type K Pipe Clamp Probe
>> Spring-Loaded Jaw for Secure Grip on Pipe (0.25" to 1.5")
>> Measurement Range: -20°C to 93°C (-4°F to 200°F)
>> Accuracy: ±1.8°C (3.6°F)
>> Use with Thermometer or Multimeter with Type K TC Input
>> Includes 3 ft Cable and Type K Connector
Part #: TP200

USB Power Adapter USB Power Adapter
>> 2.5 Watt with Interchangeable Plugs
>> Includes Plug for US

Part #: USB-POWER-ADAPTER

CompactFlash® Card Reader CompactFlash® Card Reader
>> Reads CompactFlash Cards
Part #: CF_Card_Reader

512MB CompactFlash Card 512MB CompactFlash Card
>> 1.0GB Storage Size
>> Requires CF Card Reader

Part #: 512MB_CF_CARD

512MB CompactFlash Card 1 GB CompactFlash Card
>> 1.0GB Storage Size
>> Requires CF Card Reader

Part #: 1GB_CF_CARD

512MB CompactFlash Card 2 GB CompactFlash Card
>> 1.0GB Storage Size
>> Requires CF Card Reader

Part #: 2GB_CF_Card

Specifications:

Maximum Logging Rate/Channel 2 Samples per second to PC
1 Sample per second to CF card
A/D Converters Four (4) Dual 24-bit Sigma-Delta type
Number of Channels 8 Differential Channels
Input Isolation 500 VDC minimum between field wiring and USB interface
Channel configuration Software programmable to match sensor type
Differential Input Voltage Range Thermocouple: ±0.080 Volts
RTD: 0 to 0.5 Volts
Thermistor: 0 to 2 Volts
Semiconductor Sensor: 0 to 2.5 Volts
Absolute Maximum Input Voltage Power On: ±25 Volt
Power Off: ±40 Volt
Input Impedance 5 Gigohm Minimum
Input Leakage Current Disabled Open Thermocouple Detect: 30 nA Maximum
Enabled Closed Thermocouple Detect: 105 nA Maximum
Rejection Ratio Normal Mode: 90 dB Minimum
Common Mode: 100 dB Minimum
Resolution 24 Bits
Input Coupling DC
Warm-up Time 30 Minutes Minimum
Open Thermocouple Detect Automatically enabled when the channel pair is configured for thermocouple sensors. The maximum open detection time is 3 seconds.
CJC Sensor Accuracy 15 °C to 35 °C ±0.25 °C Typical
±0.5 °C Maximum
0 °C to 70 °C -1.0 to +0.5 °C Maximum
Digital Input/Output Type CMOS
Number of Digital I/O 8 (DIO0 to DIO7)
Digital I/O Configuration Independently configured for input or output. 
Power on reset is input mode unless bit is configured for alarm.
Digital Pull up/Pull down Configuration All pins pulled up to +5 V via 47 K resistors (default). Pull down to ground (GND) also available.
Digital I/O Transfer Rate
(software paced)
Digital Input: 50 port reads or single bit reads per second
Digital Output: 100 port writes or single bit writes per second
Digital Input High Voltage High Voltage: 2.0 V Minimum; 5.5 V absolute Maximum
Low Voltage: 0.8 V Maximum; -0.5 V absolute Minimum
 
Digital Output Low Voltage (IOL = 2.5 mA): 0.7 V Maximum
High Voltage (IOH = -2.5mA): 3.8 V Minimum
Alarm Input Modes 1. Alarm when input temperature > T1
2. Alarm when input temperature > T1, reset alarm when input temperature goes below T2
3. Alarm when input temperature < T1
4. Alarm when input temperature < T1, resetalarm when input temperature goes above T2
5. Alarm when input temperature is < T1 or > T2
NOTE: T1 and T2 may be independently set for each alarm.
Alarm Output Modes Disabled: Digital I/O line may be used for normal operation.
Enabled: Active High Output (Digital I/O line goes high when a alarm conditions met)
Enabled: Active Low Output (Digital I/O line goes low when alarm conditions met)
Alarm Update Rate 1 Second
Standalone Power Supply USB Power Adapter
2.5 Watt USB Adapter with Interchangeable Plugs 
(Includes Plug for USA)
Memory Card Type Type: CompactFlash® Card 
Storage Capacity: 2GB Maximum
File System Format DOS FAT-16; the memory card must be formatted by the device prior to use for logging. The device only creates 8.3 file names and a flat directory structure ( no subdirectories). The memory card is readable by Windows operating systems after device format/use.
Logging Rate Minimum 1 second between entries, Maximum 232 seconds, 1 second granularity
Data Items Logged Timestamp, temperature or raw reading from selected channels, state of DIO lines, CJC Sensor Readings
Logging Start Methods Configurable:
1. Start Logging on Power Up - Logging begins 5 seconds after power on to allow hardware to settle.
2. Start Loggin on Button - Device is idle on power on. Press and hold button until LED comes on to begin logging. The first sample will be taken 1 second after LED comes on unless less than 5 seconds have elapsed since power on.
3. Start Logging at Specified Time - Device is idle until the real tme clock indicates the time is equal to or greater than the specified time, at which time the LED with come on. The first sample will be taken 1 second after LED comes on unless less than 5 seconds have elapsed since power on.

NOTE: Data Logging is not allowed when the device is attached to an active USB bus due to processing limitations. The device must be connected to the standalone pwer supply to perform data logging.
Logging Stop Methods Stop on button Press - To stop logging, press and hold button until LED turns off. 

NOTE: The device caches log data in volatile memory prior to writing to memory card. When logging, always use the button to stop logging and ensure data is written to memory card prior to removing power.
Logging Status Indication The LED operations when connected to the AC Adapter power supply are different than when connected to USB:

Logging modes:
1. Logging Off mode: the LED is off (disabled).
2. Start Logging on Power Up mode: the LED is on, with a momentary off flash every time data is captured.
3. Start Logging on Button mode: the LED is initially off. When the button is pressed and held for approxiamately 1 second the LED will turn on and act the same as Start Logging on Power Up mode.
4. Start Logging at Specified Time mode: the LED is off, with a momentary on flash every second until the specified date/time is reached. At that time, the LED will turn on and act the same as Start Logging on Power Up mode.

Other indication:
To Stop logging and store the remaining data to memorycard, press and hold the button until the LED turns off. It is then safe to remove the memory card.
If the memory card becomes full the LED will blink rapidly (250 ms period).
If the memory card is removed while logging is in progress the LED will blink rapidly (250 ms period). Inserting a memory card will stop the blinking.
Operating Range Temperature: 0 to 70°C
Humidity: 0 to 90% non-condensing
Storage Range Temperature: -40 to 70°C
Humidity: 0 to 90% non-condensing
OS Compatibility Windows 7, Vista, XP SP2+ (32-Bit or 64-Bit)
Dimensions 127mm x 88.9mm x 35.56mm

High Speed 8 Channel Data Logger Applications

  • Warehouses and Vehicle Cargo Spaces
  • HVAC and Refrigeration
  • Mold Control
  • Greenhouses and Nurseries
  • Wine Cellars
  • Documentation Storage