The hand held Telaire 7001 Monitor measures Carbon Dioxide and Temperature. The monitor provides stable, highly accurate readings with its patented dual beam NDIR technology. The CO2 Monitor features a 0-4 Volt output, large LCD display with push button interface, quick adjustments for altitude correction, simple toggling between °C and °F temperature readings and runs for 70+ hours on four AA alkaline batteries or continuously with the supplied power adapter. The sensor has a CO2 range suitable for virtually any building (0-10,000 ppm). Click herefor more on monitoring building air.
The CO2 Monitor is also available with a Data Logger as a Kit.
Telaire 7001 CO2 Data Logger Kit
>> (1) Telaire 7001 Carbon Dioxide Monitor (#7001)
>> (1) 4 Channel Data Loggers (#U12-012) or (#U12-013)
>> (1) Cable Connecting Logger to Monitor (#CABLE-C02)
>> (1) HOBOware Lite Software (#BWH-LITE)
The Telaire is Currently Out of Stock from the Manufacturer Until July, 2012.
NOTE: Large Quantity Discounts available!
Carbon Dioxide Monitor Features
- Monitor Carbon Dioxide and Temperature Readings
- Provides a 0 to 4 Volt Output
- 0 to 10,000 ppm Carbon Dioxide Measurement on Display
- 0-4,000 ppm Carbon Dioxide Measurement Output
- Available as a Kit with Data Logger
Description | Part Number | |||
Telaire 7001 Carbon Dioxide (CO2) Monitor | TELAIRE-7001 | |||
Telaire CO2 Kit (Includes Telaire Monitor, HOBO U12-012 Logger, Monitor to Logger Cable and HOBOware Lite Software) | 7001D-KIT | |||
Telaire CO2 Kit (Includes Telaire Monitor, HOBO U12-013 Logger, Monitor to Logger Cable and HOBOware Lite Software) | 7001D-KIT2 | |||
The CO2 Monitor is compatible with the following HOBO data loggers |
HOBO U12 4 Channel External Input Data Logger >> 4 input Channels Accepting External Sensor Probes
>> Store 43,000 Measured Readings >> 0 to 2.5 Volts DC Measurement Range >> Requires HOBOWare Software (Pro or Lite) >> Includes Mounting Kit, Hook and Loop Wire, Tie Wrap and Screws |
|
|||||||
HOBO U12 Temp/RH/Light/External Data Logger >> Temperature, Humidity, Light and 1 Analog Input Channel
>> Stores up to 43,000 Measure Readings >> Requires HOBOWare Software (Pro or Lite) >> Includes Mounting Kit w/ Magnet, Hook & Loop Tape, Tie-Wraps and Mounting Screws |
|
|||||||
HOBO U12 Temp/RH/2x External Data Logger >> Temperature, Humidity and 2 Analog Input Channels
>> Stores up to 43,000 Measured Readings >> Requires HOBOWare Software (Pro or Lite) >> Includes Mounting Kit w/ Magnet, Hook & Loop Tape, Tie-Wraps and Mounting Screws |
|
HOBO Data Logger to CO2 Monitor Cables - Required | ||||
Telaire CO2 Monitor to Data Logger Cable | CABLE-CO2 | |||
The CO2 Monitor is compatible with the following TandD data loggers | ||||
TandD Wireless Voltage/Pulse Logger | RVR-52A | |||
TandD 2 Channel Voltage Logger | VR-71 | |||
TandD Data Logger to CO2 Monitor Cables - Required | ||||
CO2 and Temperature Output Cable with 2 Leads | 2070 | |||
HOBO Data Logger Software Kits | ||||
HOBOware Lite Software Kit for Windows/Mac (includes Software and Cable) | BHW-LITE-CD | |||
HOBOware Pro Software Kit for Windows/Mac (includes Software and Cable) | BHW-PRO-CD | |||
Optional Accessories | ||||
AA Batteries (Lithium) 4 Pack | HWSB-LI |
Carbon Dioxide Monitor Specifications
Performance CO2 Channel | ||||
---|---|---|---|---|
Measurement Range | 0-4,000 ppm Voltage Output 0-10,000 ppm Display |
|||
Pressure Dependence | 0.13% of reading per mm Hg (Corrected via user input for elevation) |
|||
Sensitivity | ± 1 ppm | |||
Repeatability | ± 20 ppm | |||
Temperature Dependence | ± 0.1% of reading per °C; or ± 2 ppm per °C; whichever is greater, referenced to 25°C; |
|||
Accuracy | ± 50 ppm or 5% of reading, whichever is greater | |||
Annual Drift | ± 20 ppm typical | |||
Response Time | < 60 seconds for 90% of step change | |||
Warm-Up Time | < 60 seconds at 22°C | |||
Operating Conditions | Temperature:32°F to 122°F (0° to 50°C) Humidity: 0-95% RH, Non-Condensing |
|||
Storage Temperatures | -40°F to 140°F (-40°C to 60°C) | |||
Calibration Interval | 12 months, offset adjustment using single gas at 0-1000 ppm CO2. Factory calibration available | |||
Temperature Channel | ||||
Temperature Range | Voltage Output: 32°F to 104°F (0°C to 40°C) Display: 32°F to 122°F (0°C to 50°C) |
|||
Display Resolution | 0.1°F (0.1°C) | |||
Display Options | °F, °C, or Off. Set with Panel Button | |||
Accuracy | ± 2°F (± 1°C) | |||
Response Time | 20-30 minutes (Case must Equilibrate with Environment) |
|||
Calibration Interval | 12 months, offset adjustment using temperature standard at 50°F to 86°F (10°C to 30°C). Full factory calibration available |
|||
Other Specifications | ||||
Method | Dual Beam Absorption Infrared™ | |||
Sample Method | Diffusion or flow through (50 - 100 ml/min) | |||
Warranty | 18 months parts and labor | |||
Output | ||||
Analog | CO2: 0-4 VDC, 1mV/ppm (4,000 ppm max) Temperature: 0-4 VDC linear, 32°F to 104°F (0°C to 40°C) |
|||
Output Impedance | 100 Ohms | |||
Wiring | Connection Via RJ-45 to DB9 Serial Port Cable | |||
Display | LCD with independent CO2/ temp readings (panel buttons set elevation, °F/°C, calibration functions) | |||
Power Supply | ||||
Battery Type | Four AA Alkaline Batteries (Not Included) | |||
Battery Operation | 80 Hours (alkaline) | |||
External | 6 VDC from external AC/DC adapter (included) | |||
Power Requirements | 100 mA Peak, 20 mA average from 6V Certification |